Application, rating and electric data
| Item | Specification |
| Application | Storage of general (non-flammable) laboratory materials |
| Storage Volume | 25 liters (0.9 cu.ft.) |
| Storage Capacity | 18 standard 2” boxes |
| Temperature Range | -86°C to -20°C @ 32°C (90°F) ambient, uniformity +/- 3°C at -80°C top to bottom, adjustable in 1°C increments. Default temperature setting is -80°C. Preset options are -86°C, -40°C and -20°C |
| Electric Power | 100V to 240VAC at either 50 or 60Hz or 12V DC from mobile source |
| Power Cord Options Available | Multiple options are available. Consult with your sales representative![]() |
| Maximum Power (Current) | 280 watts (4 amps @120VAC, 2 amps @240VAC, 15 amps @ 12V DC) |
| Auto-Voltage Capability | 100V to 240V, 50 or 60 Hz |
| Electric Supply Rating | 15 amp or greater grounded circuit |
| Certification/Agency Listing | CE, cTÜVus |
| Noise | Advanced noise abatement, <45 dB(A) at 1 meter |
| Indoor/Outdoor Use | Indoor use only |
| Environmental Conditions | Non-corrosive, non-flammable, non-explosive, indoor use, altitude up to 2,000m, temperature 5°C to 40°C (41°F to 104°F), maximum relative humidity 80% for temperature up to 31°C decreasing linearly to 50% at 40°C |
Controller
| Item | Specification |
| Controller Type | Microprocessor controls |
| Security | Lockable lid |
| Warm and Cold Alarms | Setpoint ± 10°C |
| Control Sensor | One RTD (PT100 Class A) |
| Internet Connectivity | Optional, independent wireless temperature monitoring and logging |
Refrigeration system
| Item | Specification |
| Cooling Engine | Helium charged free-piston Stirling engine with continuous modulation |
| Heat Transport System | Gravity driven thermosiphon |
| Refrigerant | R-170 (Ethane) 10–12 grams |
| Evaporator | Cold wall (inner liner) |
| Heat Rejection | Finned heat exchanger with forced air cooling Air inlet: Right front side of unit Air outlet: Below control panel of front mechanical compartment |
| Defrost Method | Manual |
Performance data
| Item | Specification |
| Steady State Energy Use at 25°C Ambient | 2.8 kWh/day (average power 118 watts) at -80°C (Empty Cabinet) |
| Pull-Down from 25°C Ambient | 4 hours to -80°C (Empty Cabinet) |
| Recovery from 1 minute lid opening | 20 minutes to -80°C (Empty Cabinet) |
| Warm-up Profile | 30 minutes to -60°C (Empty Cabinet) 70 minutes to -40°C (Empty Cabinet) |
| Heat Dissipation | 403 BTU/h (load to HVAC) (Empty Cabinet) |
Dimensions and construction
| Item | Specification |
| Interior (L x W x D) | 332 x 221 x 340 mm / (13.1 x 8.7 x 13.4 in.) |
| Exterior (L x W x D) | 692.5 x 350 x 460 mm / (27.3 x 13.8 x 18.1 in.) |
| Net Weight, Empty | 21 kg (46 lbs.), nominal |
| Shipping (L x W x D) | 762 x 457 x 559 mm / (30 x 18 x 22 in.) |
| Shipping Weight | 24 kg (53 lbs.) |
| Insulation | High performance vacuum insulated panels and polyurethane foam |
| Gasket Heater | Continuous operation when the power cord is connected |
| Included Accessories | Stacking utility bin (x 2), insulated sub-lid, AC and DC power cords, removable port seal |
| Options | Certificate of Calibration, Stirling Ultraguard -70C phase change material |